- Inconel™ 600 sheath that’s 305 mm (12 in) long and 6.35 mm (0.250 in) in diameter - Calibrated accuracy of ± 0.012 °C at 0 °C. - Reference-grade platinum sensing element. - NVLAP-accredited calibration, lab code 200706-0. - Durable sensors, ideal in the factory, field, or lab.
Index
วัดอุณหภูมิ
Temperature range
5615-12 and 5615-9: –200 °C to 420 °C
Accuracy[1]
± 0.024 °C at –200 °C
± 0.012 °C at 0 °C
± 0.035 °C at 420 °CC
Short-term repeatability[2] ± 0.009 °C at 0.010 °C
Drift[3] ± 0.007 °C at 0.010 °C
Transition junction dimensions 71 mm x 13 mm dia (2.8 in x 0.5 in)
Sheath dimensions 305 mm x 6.35 mm (12.0 in x 0.250 in)
Sensor length 254 mm (10 in)
Sensor location 6.9 mm ± 3.3 mm from tip (0.27 in ± 0.13 in)
Fluke-5615-12-P X1
Sheath diameter tolerance ± 0.127 mm (± 0.005 in)
Sheath material Inconel™ 600
Minimum insulation resistance 1000 MΩ at 23 °C
Transition junction temperature range[4] –50 °C to 200 °C
Self heating (in 0 °C bath) 50 mW/°C
Lead-wire cable type PTFE insulated with PTFE jacket, 22 AWG
Lead-wire length 183 cm (72 in or 6 ft)
Lead-wire temperature range –50 °C to 200 °C
Calibration
NVLAP-accredited calibration included, lab code 200706-0. Please see calibration uncertainty table and its explanation of changeable uncertainties.
Referance
[1] Includes calibration and 100 hr drift (k=2)
[2] Three thermal cycles from min to max temp, includes hysteresis, 95 % confidence (k=2)
[3] After 100 hrs at max temp, 95 % confidence (k=2)
[4] Temperatures outside this range will cause irreparable damage. For best performance, transition junction should not be too hot to touch.
[5] Per ASTM E 644