- Designed for high-temperature applications, up to 661°C. - Meet all ITS-90 requirements for resistance ratios. - Exhibit an Rtp drift of less than 20 mK after 500 hours at 661°C. - Calibrated accuracy of ± 0.006 °C at 0 °C.
5628-15-P ฟังก์ชั่นการวัด
วัดอุณหภูมิ
5628-15-P ช่วงการวัด
Temperature Range 25.5 Ω, –200 °C to 661 °C
RTPW 100 Ω (± 1 Ω)
Resistance Ratio W(Ga) W(302.9146K) 1.11807 α 0.003925
Calibrated Accuracy† (k=2) ± 0.006 °C at –200 °C // ± 0.006 °C at 0 °C // ± 0.015 °C at 420 °C // ± 0.022 °C at 661 °C
Stability ± 0.003 °C
5628-15-P ขนาดสินค้า
6.35 mm dia. x 305 mm, 381 mm, or 508 mm (0.25 x 12, 15, or 20 in) standard, custom lengths available
5628-15-P อุปกรณ์ที่มาในชุด
โพรบ 1 เส้น
5628-15-P ข้อมูลจำเพาะ
Long-Term Drift (k=2) 5626: < 0.006 °C/100 hours at 661 °C
Immersion At least 12.7 cm (5 in) recommended
Sheath Inconel™ 600
Lead Wires 4-wire Super-Flex PVC, 22 AGW
Termination Gold-plated spade lugs, or specify
Calibration Accredited calibrations from Fluke Calibration